EFAI Chestsuite XR Malpositioned ETT Assessment System (ETT-XR-100)

Ever Fortune.AI Co., Ltd.
Product Details
Model Identifier
Manufacturer
Ever Fortune.AI Co., Ltd.
Product
EFAI Chestsuite XR Malpositioned ETT Assessment System (ETT-XR-100)
Version
Unknown
Date Cleared
02/20/2025
FDA Submission No.
Category
CADt
Model Characteristics
Inclusion Criteria
Chest xray images of adults
Exclusion Criteria
Unknown
Instructions for Use
Not available
Indications for Use
Indication of Use
EFAI CHESTSUITE XR MALPOSITIONED ETT ASSESSMENT SYSTEM (EFAI ETTXR) is a radiological computeraided triage and notification software indicated for use in the analysis of chest X-ray (CXR) images in adults. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of vertically malpositioned endotracheal tube (ETT) in relation to the carina. Findings are flagged when the ETT distal tip is assessed as being more than 7 cm above the carina, less than 3 cm above the carina, or when it is below the carina (i.e in the right or left mainstem bronchus). The device assesses solely the vertical position of the ETT distal tip relative to the carina, does not factor patient positioning, and cannot detect esophageal intubation. The device is tested in the single lumen endotracheal tube, while it may trigger a false prioritization alert in the case of properly positioned double lumen ETT. EFAI ETTXR analyzes cases using deep learning algorithms to identify suspected malpositioned ETT findings. It makes case-level output available to a PACS/workstation for worklist prioritization or triage. EFAI ETTXR is not intended to direct attention to specific portions of an image or to anomalies of an image. Its results are not intended to be used on a stand-alone basis for clinical decision-making nor is it intended to rule out malpositioned ETT or otherwise preclude clinical assessment of chest radiographs.
Intended User
Trained medical specialists
Age
Adult
Anatomy
Chest
Modality
XRAY
Output
Malpositioned Endotracheal Tube (ETT) (Triage)
Details on Training Data Sets
Details on Training Data Sets
No. of Cases
Unknown
Age Range (Years)
Unknown
Sex (%)
  • Female: Unknown
  • Male: Unknown
  • Unknown: Unknown
Output
Unknown
Race (%)
  • White: Unknown
  • Black or African American: Unknown
  • American Indian or Alaska Native: Unknown
  • Asian: Unknown
  • Native Hawaiian or Other Pacific Islander: Unknown
  • Unknown: Unknown
Ethnicity (%)
  • Hispanic or Latino: Unknown
  • Not Hispanic or Latino: Unknown
  • Unknown: Unknown
Geographic Region (%)
  • USA: Unknown
  • International: Unknown
  • Unknown: Unknown
Scanner Manufacturer(s)
Unknown
Scanner Model(s)
Unknown
Model Performance
Study Type
Performance Testing Type
Stand-Alone Performance
Standalone Model Performance
Reference Standard (Ground Truth)
Interpretation by Reviewing Clinician
No. of Cases
940
Age Range (Years)
Unknown
Sex (%)
Female: 49
Male: 50
Race (%)
  • White: Unknown
  • Black or African American: Unknown
  • American Indian or Alaska Native: Unknown
  • Asian: Unknown
  • Native Hawaiian or Other Pacific Islander: Unknown
  • Unknown: Unknown
Ethnicity (%)
  • Hispanic or Latino: Unknown
  • Not Hispanic or Latino: Unknown
  • Unknown: Unknown
Geographic Region (%)
  • USA: Unknown
  • International: Unknown
  • Unknown: Unknown
Output
Positive: 28% Negative: 72%
Scanner Manufacturer(s)
Philips;Samsung;Siemens;Canon;Carestream
Scanner Model(s)
Unknown
No. of Sites
Unknown
Model Accuracy
Unknown
Model Sensitivity
89% (95% CI: 84
Model Specificity
93
Reader Study Performance
No. of Readers
N/A
No. of Cases
N/A
No. of Sites
N/A
Output
N/A
Age Range (Years)
N/A
Sex (%)
  • Female: N/A
  • Male: N/A
  • Unknown: N/A
Race (%)
  • White: N/A
  • Black or African American: N/A
  • American Indian or Alaska Native: N/A
  • Asian: N/A
  • Native Hawaiian or Other Pacific Islander: N/A
  • Unknown: N/A
Ethnicity (%)
  • Hispanic or Latino: N/A
  • Not Hispanic or Latino: N/A
  • Unknown: N/A
Geographic Region (%)
  • USA: N/A
  • International: N/A
  • Unknown: N/A
Scanner Manufacturer(s)
N/A
Scanner Model(s)
N/A
Model Accuracy
N/A
Model Sensitivity
N/A
Model Specificity
N/A
Model Limitations, Warnings, & Precautions
Model Limitations, Warnings, & Precautions
Supported Scanner Manufacturer(s)
Unknown
Slice Thickness
Unknown
Contrast Use
Unknown
MRI Field Strength
Unknown
Reconstruction Kernel Used
Unknown
Alternative Choices
Alternative Choices
Previous Version(s)
Unknown
Predicate Device(s)
Unknown
Contact Information
Contact Information
Point of Contact Name
Unknown
Email
Unknown
Additional Details
Related Use Cases
Unknown
All Products And Platforms from Ever Fortune.AI Co., Ltd.